Ecoclave S & B

  • Print

Ortoalresa’s ecoclaves class S y B are versatile devices easy to use. They have 12 programs on memory pre-fixed in factory for a smooth processing. They include: trays holder, 4 trays, door key/clamp, 2 hoses, sponge and funnel. Include standard printer.

 

Features

  • Volume 18 & 23 litres.
  • Continuos control over the cycle phases.
  • Instantaneous vaporizer inside the chamber.
  • Provided with connection to demineralizer
  • Humidity <2%.

User friendly

  • Ecoclave S: 2 Test Cycles: Bowie & Dick, Vacuum Test ,Pre and Post Vacuum.
  • Ecoclave B: 3 Test Cycles: Bowie & Dick, Hellix test, Vacuum Test.
  • Cycles with clean water full tank: Class S: 10 for 18 l, 8 for 23 l. Class B: 8 for 18 l and 6 for 23 l.
  • Weight of wrapped materials 18 l 3 Kg, 23 l 4,5 Kg. Unwrapped: 18 l 4 Kg, 23 l 6 Kg.
  • Porous: 18 l 1Kg (B) /0,5 Kg (S) ,23 l 1,5 Kg (B)/1 Kg (S).
  • Multi-language graphic display with temperature, Sterilization and drying time, chamber pressure, program and stage of cycle.
  • Night cycle.
  • Initial checking and tracking through messages.
  • Stainless steel chamber made of thick one-piece molded steel.
  • Loading with self-priming pump.
  • Electronic drying in vacuum
  • Memory Test System (MTS): store and print 10 cycles.

Safety

  • Electrical: fuses and earthed.
  • Motor-operated closure with triple-protection,
  • Safety valve.
  • Several messages keep the operator informed about the device situation.
  • Air entry through bacteriological filter.
  • Double-stage vacuum pump release the air avoiding air bubbles.

Directives

EU Directives: 98/79/EC, 2006/95/EC, 2004/108/EC, 2011/65/EU, 2012/19/EU.
Standards: EN 61010-1, EN 61010-2, EN 61326-1, EN 13060.

Versions

Ref Dimensions (mm)
(w x d x h)
Net weight (Kg) Trays dimensions(mm) Class Power
(W)
Chamber volume (l) Test cycle
AU 005 505 610 400 56 185x285 B 2400 18 B+V+H
AU 006 505 695 400 60 185x440 B 2400 23 B+V+H
AU 007 450 610 400 45 185x285 S 2400 18 B+V
AU 008 450 695 400 50 185x440 S 2400 23 B+V

ATENTION: for 23 liter models drying time lasts 5 minutes, total time 15 min.
THIS AUTOCLAVE CAN NOT STERILIZE LIQUIDS.
B=Bowie & Dick | V= Vacuum Test | H= Helix

Programs

Instruments & materials Cycle Ecoclave B Ecoclave S Wrapped
Total time (min) No. vaccum Total time (min) No. vaccum
Delicate holow and stainless steel 121˚C hollow wrapped 26 4 30 2 YES
Stainless steel hollow 134˚ Hollow wrapped 14 4 22 2 YES
Solid rubber /delicate 121˚Solid wrapped 26 2 28 2 YES
Stainless steel solid 134˚ Solid wrapped 14 2 15 2 YES
Solid and stainless steel hollow 134˚ Prion 30 4 35 2 YES
Delicated, hollow and porous 121˚Porous 31 4 33 2 YES
Solid stainless steel, hollow, small aorous 134˚ Porous 19 4 20 2 YES
Soldid rubber and delicate 121˚Rapid 20 2 23 2 NO
Solid stainless steel 134˚Rapid 8 2 10 2 NO
Hollow rubber and stainless steel 134˚ Hollow unwrapped 8 4 11 2 NO
Ciclo Test Helix /Bowie &Dick 134˚ Helix/B&D Test 7,5 4 -/7.5 -/2 -
Test cycle vacuum <40˚ Vacuum Test 15 1 15 1 -