Digtor 21

The centrifuge Digtor 21 is a versatile device, allowing be present on routine labs, research, biotechnology, quality control, hospitals, etc.

It is provided of several advantages as:
Innovation: with a TFT touch screen allowing operator the highest autonomy,
controlling the separation, as well as a save energy system selectable.
Versatility: since 4 x 750 ml, 52 conical tubes of 15 ml, rotor for microplates, microtubes,
and a wide range of adapters.
Safety: always shown on screen real values of RCF, based on accessories selection,
automatically, no data introduction is required. Accessories for bio containment.
Efficiency: Max RCF 20.804 xg, selection of acceleration and braking time, depending
of sample performances.

Your best partner on lab.

 

User friendliness

  • TFT touch screen easy to read and values selection (RPM on steps of 100, RCF on
  • steps of 10xg), time, temperature (1°C steps), acceleration/deceleration (1 second steps).
  • The screen which indicates the status can be seen from more than 3 m.
  • Rotor and adapters list on memory.
  • Real values on screen of RCF based on accessories configuration.
  • Timer count up/down, from 0 or at “set RPM/RCF” for test reproducibility.
  • Short spin key with speed adjustable.
  • Progressive acceleration and braking selectable by operator up to 175 ramps.
  • Program data protection through password selectable.
  • Automatic open lid (only non-refrigerated).
  • Automatic rotor recognition.
  • Automatic lid latch lock.
  • Microprocessor controlled.
  • Option: free or locked adjustment of RPM/RCF along the run.

Features

  • Induction motor maintenance free
  • High capacity: Max. Volume 4 x 750 ml.
  • High speed, up to 20.804 xg/14.300 RPM.
  • Quiet: noise level < 60 dB.
  • 16 memories for programs.

Safety

  • Lid locking and holding.
    Imbalance switch off.
    Manual overriding.
    Lid dropping protection.
    Chamber of centrifugation in stainless steel.
    Viewing and calibration prot in the lid.
    Several messages keep the operator informed about the device situation.
    Guard ring between the chamber of centrifugation and the housing.
    Forced ventilation which reduces the temperature increase (only non-refrigerated).

Standards and EC Directives

EU Directives: 98/79/EC, 2006/95/EC, 2004/108/EC, 2011/65/EU, 2012/19/EU.
Standards: EN 61010-1, EN 61010-2-020, EN 61326-1.

 

Versions

Code Dimensions (mm)(w x d x h) Net weight (Kg) Voltage
(V)
Frecuency
(Hz)
Consumption
(W)
CE 112 540 650 390 50 220-240 50-60 980
CE 118 540 650 390 50 110-120 50-60 980

accesorios

  RT 191 (1)
id
RT 192
id
RT 195
id
RT 238
id
RT 197
id
  + info + info + info + info + info
ROTOR SWING OUT SWING OUT SWING OUT ANGLE FIXED 30º ANGLE FIXED 45º
Max. capacity 4x750 ml. 4x250 ml. 104x5ml. 6x85 ml. 8x50 ml.
RPM 3700 4200 3800 9000 6000
Radius (mm) 204 202 185 112 149
RCF Max. (xg) 3122 3984 2987 10142 5997
Min. temp. max speed (°C) 0 1 0 1 0
  RT 198
id
RT 199
id
RT 201 (1)(2)
id
RT 202
id
RT 206
id
  + info + info + info + info + info
ROTOR ANGLE FIXED 45º ANGLE FIXED 30º SWING OUT SWING OUT ANGLE FIXED 30º
Max. capacity 4x100 ml. 4x250 ml. 4 bolsas sangre 12 p. microtiter 30x1,5-2 ml.
RPM 5600 4700 3700 3700 14300
Radius (mm) 138 153 204 182 (3) 96
RCF Max. (xg) 4838 3779 3122 2786 21948
Min. temp. max speed (°C) -1 -4 0 -4 -1

(1) Availables lids for that buckets (RE 356)
(2) This rotor can fit adapters for blood bags (RE 308).
(3) Medium radius on bucket.